Int J Stomatol ›› 2022, Vol. 49 ›› Issue (3): 324-327.doi: 10.7518/gjkq.2022052

• Original Articles • Previous Articles     Next Articles

Exploration on cleaning and disinfection mode of dental instruments based on multi-enzyme detergent control

Lin Jie1(),Liu Fan2()   

  1. 1.State Key Laboratory of Oral Diseases & National Clinical Research Center for Oral Diseases & Dept. of Implantology, West China Hospital of Stomatology, Sichuan University, Chengdu 610041, China
    2.State Key Laboratory of Oral Diseases & National Clinical Research Center for Oral Diseases & Dept. of Nursing, West China Hospital of Stomatology, Sichuan University, Chengdu 610041, China
  • Received:2021-09-20 Revised:2022-01-10 Online:2022-05-01 Published:2022-05-09
  • Contact: Fan Liu E-mail:84204362@qq.com;samotj@163.com
  • Supported by:
    Horizontal Project of Sichuan University(21H0441)

Abstract: Objective

To obtain the optimal temperature, the operating life after the configuration of the detergents, and the best reuse times of the detergents corresponding to the two kinds of multi-enzyme detergents, which were commonly used in the cleaning and disinfection of dental instruments.

Methods

In this study, the multi-enzyme detergents currently in commercial use were taken as the research objects, according to the best working temperature, configuration time and frequency of fluid change, explore the optimal conditions for the use of multi-enzyme cleaning agents.

Results

The optimal temperature of group one treatment is around 55 ℃, while the optimal temperature of group two is around 50 ℃. Too long detergent configuration time or repeated use of multiple enzyme detergent in batches may result in reduced disinfection effect. Reasonable control of the number of instruments, use within an effective time period, and timely replacement of multi-enzyme detergents are important methods to ensure the disinfection effect.

Conclusion

Controlling the working temperature at about 50-55 ℃, using the multi-enzyme detergents of the configuration time, and the cleaning times of not more than three times can help to clean the dental instruments efficiently.

Key words: multi-enzyme detergent, oral instrument, hospital infection control

CLC Number: 

  • R 187

TrendMD: 

Fig 1

Effect of temperature on cleaning time of different multi enzyme cleaning agents"

Fig 2

Effect of configuration time on cleaning time of different multi enzyme cleaning agents"

Fig 3

Effect of cleaning batch on cleaning time of different multi enzyme cleaning agents"

1 陈继峰, 卞莲莲, 刘洪源, 等. 医用多酶洗涤剂洗涤效果评价方法探讨[J]. 中国卫生检验杂志, 2013,23(4): 877-879.
Chen JF, Bian LL, Liu HX, et al. Methods for determining the ablution effect of multi-enzyme detergents for hospital application[J]. Chin J Health Lab Technol, 2013, 23(4): 877-879.
2 Hadi R, Vickery K, Deva A, et al. Biofilm removal by medical device cleaners: comparison of two bioreactor detection assays[J]. J Hosp Infect, 2010, 74(2): 160-167.
3 Stiefel P, Mauerhofer S, Schneider J, et al. Enzymes enhance biofilm removal efficiency of cleaners[J]. Antimicrob Agents Chemother, 2016, 60(6): 3647-3652.
4 Lawson VA, Stewart JD, Masters CL. Enzymatic detergent treatment protocol that reduces protease-resistant prion protein load and infectivity from surgical-steel monofilaments contaminated with a human-derived prion strain[J]. J Gen Virol, 2007, 88(10): 2905-2914.
5 张爱勤, 刘琴, 王晓丽, 等. 多酶洗液加超声清洗器清洗医疗器械效果[J]. 中国消毒学杂志, 2008, 25(5): 556-557.
Zhang AQ, Liu Q, Wang XL, et al. Effect of multi-enzyme lotion and ultrasonic cleaner on cleaning medical instruments[J]. Chin J Disinfect, 2008, 25(5): 556-557.
6 石艳燕. 多酶清洁剂与数控超声波在清洗口腔小器械中的应用[J]. 中华医院感染学杂志, 2011, 21(19): 4173.
Shi YY. Application of multi-enzyme cleanser and numerical control ultrasonic in cleaning oral instruments[J]. Chin J Nosocomiol, 2011, 21(19): 4173.
7 何渺. 超声震荡清洗和传统人工冲刷对车针清洗效果的比较[J]. 山西医药杂志, 2013, 42(9): 1076.
He M. Comparison between ultrasonic shock cleaning and traditional manual washing on cleaning effect of car needle[J]. Shanxi Med J, 2013, 42(9): 1076.
8 罗万军, 徐润琳, 王文娟, 等. 污染器械预处理方式及放置时间对清洗效果的影响研究[J]. 中国消毒学杂志, 2015, 32(12): 1188-1190.
Luo WJ, Xu RL, Wang WJ, et al. Investigation on cleaning effect to contaminated instruments after different pretreatment and storage time[J]. Chin J Disinfect, 2015, 32(12): 1188-1190.
No related articles found!
Viewed
Full text


Abstract

Cited

  Shared   
  Discussed   
[1] . [J]. Foreign Med Sci: Stomatol, 1999, 26(06): .
[2] . [J]. Foreign Med Sci: Stomatol, 1999, 26(06): .
[3] . [J]. Foreign Med Sci: Stomatol, 1999, 26(06): .
[4] . [J]. Foreign Med Sci: Stomatol, 1999, 26(06): .
[5] . [J]. Foreign Med Sci: Stomatol, 1999, 26(05): .
[6] . [J]. Foreign Med Sci: Stomatol, 1999, 26(05): .
[7] . [J]. Foreign Med Sci: Stomatol, 1999, 26(04): .
[8] . [J]. Foreign Med Sci: Stomatol, 1999, 26(04): .
[9] . [J]. Foreign Med Sci: Stomatol, 2004, 31(02): 126 -128 .
[10] . [J]. Inter J Stomatol, 2008, 35(S1): .